5 matches found
CVE-2020-16216
CVE-2020-16216 affects Philips IntelliVue patient monitors (MX100, MX400-550, MX600, MX700, MX750, MX800, MX850, MP2-MP90) and IntelliVue X2/X3 prior to N. The issue is improper input validation in multiple components, leading to a denial-of-service via a system restart when processing input data...
CVE-2018-10597
CVE-2018-10597 affects Philips IntelliVue MP Series, MX monitors and Avalon FM monitors. Root cause: improper authentication allowing an unauthenticated attacker to read/write memory via an attacker-chosen device address within the same subnet (write-what-where). Affected versions include MP Seri...
CVE-2018-10601
The CVE-2018-10601 issue affects Philips IntelliVue Patient Monitors MP Series (and MP2/X2/MP30/MP50/MP70/NP90/MX700/800), MX (MX400-550, Rev J-M; X3/MX100 Rev M), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50. Root cause: a vulnerability in an exposed "echo" service where an attacker-se...
CVE-2020-16228
CVE-2020-16228 affects Philips PICiX (PIC iX) Versions C.02/C.03, PerformanceBridge Focal Point A.01, IntelliVue monitors MX100/MX400-MX850 and MP2-MP90, and IntelliVue X3 Versions N and prior. The issue is improper or missing certificate revocation checking, which may cause the device to trust a...
CVE-2018-10599
The Connected document (ICSMA-18-156-01) confirms CVE-2018-10599 affects Philips IntelliVue Patient Monitors (MP series, MX series) and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50. Root cause: information exposure vulnerability allowing an unauthenticated attacker to read memory from an at...